Fork me on GitHub

Project Notes

#372 MilliVoltmeterDIY

An ATmega328-based millivolt meter based on a design by Scullcom Hobby Electronics. Uses an LTC2400 ADC and LT1019 voltage reference.

See CustomBoardAndEnclosure for details of the custom PCB and enclosure I made for the project.

Build

Notes

This project is based on the excellent build tutorial by Scullcom Hobby Electronics #44 - Millivolt Meter MK2

clip

A few changes I’ve made:

  • LT1019 2.5V voltage reference instead of the ADR4540BRZ
  • direct LCD control instead of I²C (because it is what I had available)
  • some of the buttons moved to analog pins (digital mode) since the LCD uses a few too many pins
  • adjust the controls to a three-button command set.
  • discarded the op-amp buffer, as I found it skewed the reading at the top and bottom of the range
  • altered the calibration method, so the only fixed value in the code is the reference voltage.

Button Functions

I used normally-open pushbuttons on the breadboard, but for the final build I have some normally-closed pushbuttons. The #define BUTTON_LOGIC_ACTIVE_STATE is used to select between HIGH or LOW button logic. Internal pull-ups are enabled so there are no additional external components required other than the push-buttons themselves.

All the necessary functions are implemented with a three-button command set:

  • button 1: calibrate
  • button 2: change the number of digits (precision)
  • button 3: change the second-line function between 3 modes:
    • bar graph
    • hold value
    • raw ADC output

Buffering the Input Voltage

The input signal is tapped from a voltage-divider to provide around 10x attenuation, and therefore support a wider voltage range.

The values used in the voltage-divider are not critical. They just need to be in a suitable proportion to scale back the intended input voltage range to a maximum limit of the reference voltage, and large enough to provide a high-impedence input to the voltage under test.

In my design, I’m using 100kΩ and 1MΩ which support a full-range of around 27.5V

Ideally the reading should be buffered to further reduce coupling with the circuit-under-test, as in the original Scullcom project. I tested buffering with an LMC6482 op-amp, but although it is “rail-to-rail” I found it performed quite poorly at the top and bottom of the range e.g. 0V input still producted a 40mV output on the buffer. I decided to discard the buffer, and just rely on a stiff voltage divider. Note that the Scullcom project used a AD8628ARZ, so perhaps that performs better as a buffer (I will have to get some to try).

I also planned to add a 1N4733 5.1V zener diode to provide some over-voltage input protection (across the lower half of the voltage-divider). However the leakage current introduced a minor non-linearity in the readings. So again, I discarded this idea in favour of the most accurate reading possible.

Using the LTC2400

10kΩ pull-ups are added on the CS and SCK pins to ensure the chip does not go into internal oscillator mode on startup.

The LTC2400 provides a 24-bit measurement with respect to the reference voltage (Vref = x00FFFFFF). This is received in a 32-bit serial stream:

  • 4 high bits provide status
  • 4 low bits are “sub LSB” and can be discarded without affecting accuracy
  • the 24 bits in between are the actual reading

The status bits help decode the reading correctly:

  • SIG indicates if reading is positive or negative
  • EXR indicates if the reading exceeds the reference range (above or below)

Voltage Reference and Calibration

I’m using an LT1019 2.5V reference (because it is what I had available). This means that the LTC2400 ADC output is all calculated with respect to the 2.5V standard.

I have adapted the code so that the unit performs an automatic calibration with respect to the reference voltage. The only hardcoded value is the reference voltage. How this works:

  • the reference voltage is exposed on a “calibration” test point
  • the +ve input is connected to the “calibration” test point
  • press the “calibration” botton
  • the calibration adjustment factor is calculated to ensure an accurate reading of the reference voltage
  • the calibration adjustment factor is also stored in EEPROM as the default on startup

As a result - as long as the reference voltage is accurate(!) - this approach is agnostic with regards to any input attenutation set in the input voltage divider.

How Calibration and Voltage Conversion Works

The ADC measures the voltage at the split-point of the voltage divider.

When the 2.5V reference voltage is applied to the +ve input, the ADC reading thus corresponds to “2.5V”.

The code assumes a stright-line relationship with two data points:

  • 0V = 0 from the ADC
  • 2.5V = the calibrated value from the ADC

Thus given any ADC reading, the voltage = 2.5 * reading / calibration_reading

Test Results

So how well does it perform? Here are some measurements.

  • calibration factor: 0x19518F (= 2.5V)
  • Note: the DMM used here reads the 2.5V test point as 2.49V
Power Supply Vin DMM mVm ADC ADC Check Calc
0.00 0.00 27mV 0x4666 27mV
0.99 1.00 1.028V 0xA69F0 1.028
1.99 2.00 2.02V 0x1478CC 2.02
2.99 3.00 3.01V 0x1E81FC 3.01
3.99 4.00 4.00V 0x2886B0 4.00
4.99 5.00 4.99V 0x329080 4.99
5.99 5.99 5.98V 0x3C87A0 5.98
6.98 6.98 6.95V 0x466E20 6.95
7.98 7.98 7.95V 0x5077C0 7.95
8.99 8.98 8.94V 0x5A86C0 8.94
10.00 9.98 9.93V 0x6494C0 9.93
11.00 10.98 10.9V 0x6E9E30 10.92
11.99 11.97 11.90V 0x788BF0 11.90

Note:

  • “Power Supply Vin” is what the power supply said it was delivering
  • “DMM” is the reading with a digital multimeter
  • “mVm” is the reading with the MilliVoltmeter
  • “ADC” is the LTC2400 reading used by the MilliVoltmeter
  • “ADC Check Calc” is confirmation of the value reported by the MilliVoltmeter based on the ADC value

So, hmm, there appears to be some skew in the readings obtained from the LTC2400:

  • non-zero offset (0x4666)
  • skew slightly high below 4V
  • skewing slightly low above 4V

While that suggests a need to mathematically fit to a curve with a non-zero offset, that’s not how I expect the LTC2400 to behave.

Am I using the LTC2400 correctly? Re-reading the datasheet I see that while Analog Input Voltage is rated for –0.3V to (VCC + 0.3V), it is qualified by notes, in particular:

Note 14: For reference voltage values VREF > 2.5V the extended input of –0.125 • VREF to 1.125 • VREF is limited by the absolute maximum rating of the Analog Input Voltage pin (Pin 3). For 2.5V < VREF ≤ 0.267V + 0.89 • VCC the input voltage range is – 0.3V to 1.125 • VREF. For 0.267V + 0.89 • VCC < VREF ≤ VCC the input voltage range is – 0.3V to VCC + 0.3V.

Since I have VREF = 2.5V and VCC = 5.0V:

  • 0.267V + 0.89 • VCC = 4.717V
  • so my input voltage range is –0.3V to 2.8125V

With the 1MΩ : 100kΩ voltage divider on the analog input, the input voltages are actually staying within this band for the tests above: ranging from 0V to 1.091V @ 12V test voltage.

Next step: read the application info in the datasheet more carefully. See LEAP#448 LTC2400 for more on the investigation into getting good results from the LTC2400.

Code

MilliVoltmeterDIY.ino is my version of the code, based on the original Millivolt_Meter_MK2_ver33.ino.

Libraries used:

Breadboard Construction

I initially built this on a breadboard and controlled with an Arduino Uno to verify and refine the circuit.

Breadboard

Schematic

Build

Credits and References

About LEAP#372 ATmegaADCToolsTest Equipment

This page is a web-friendly rendering of my project notes shared in the LEAP GitHub repository.

Project Source on GitHub Return to the LEAP Catalog
About LEAP

LEAP is my personal collection of electronics projects - usually involving an Arduino or other microprocessor in one way or another. Some are full-blown projects, while many are trivial breadboard experiments, intended to learn and explore something interesting.

Projects are often inspired by things found wild on the net, or ideas from the many great electronics podcasts and YouTube channels. Feel free to borrow liberally, and if you spot any issues do let me know or send a pull-request.

NOTE: For a while I included various scale modelling projects here too, but I've now split them off into a new repository: check out LittleModelArt if you are looking for these projects.

Project Gallery view the projects as an image gallery Notebook reference materials and other notes Follow the Blog follow projects and notes as they are published in your favourite feed reader